Microwave-assisted Extraction Followed by Capillary Electrophoresis-Amperometric Detection for the Determination of Antioxidant Constituents in Folium Eriobotryae

Zhi Chen,Luyan Zhang,Gang Chen
DOI: https://doi.org/10.1016/j.chroma.2008.04.032
2008-01-01
Abstract:A method based on capillary electrophoresis-amperometric detection has been developed for the determination of catechin, rutin, kaempferol, gentistic acid, and quercetin in Folium Eriobotryae with the aid of microwave-assisted solvent extraction. The effects of the acidity and the concentration of the running buffer, separation voltage, injection time, detection potential, and irradiation time were investigated to acquire the optimum analysis conditions. The detection electrode was a 300-μm-diameter carbon disc electrode at a detection potential of +0.90V. The five analytes could be well separated within 12min in a 40-cm length fused-silica capillary at a separation voltage of 12kV in a 50-mM borate buffer (pH 9.0). The relation between peak current and analyte concentration was linear over about three orders of magnitude with detection limits (S/N=3) ranging from 2.1 to 4.5μM for all the analytes.
What problem does this paper attempt to address?