Preliminary Study of the Repeatability and Agreement of Scheimpflug Noncontact Tonometry (corvis ST)

Pei-pei ZU,Ya WANG,Tong ZUO,Di WU
DOI: https://doi.org/10.3760/cma.j.issn.1674-845x.2013.05.002
2013-01-01
Abstract:Objective To assess the repeatability and agreement of Scheimpflug Noncontact Tonometry (Corvis ST) on normal subjects.Methods Forty-seven myopic subjects (right eye) were enrolled in this prospective study with Scheimpflug Noncontact Tonometry (Corvis ST).The right eyes of 34 myopic subjects were measured by the Ocular Response Analyser (ORA),non-contanct tonometer (NCT) and Pentacam.The repeatability of data was analyzed by intraclass correlation coefficient (ICC) and Cronbach's Alpha coefficients,and the data were compared using a paired t test,the Pearson correlation test and Bland-Altman plot.Results The ICC and Cronbach's Alpha coefficients of 1st A-time,2nd A-time,Vout,HC-DA,IOPst and CCTst were more than 0.8,which showed high intraobserver repeatability.The Bland-Ahman plots displayed relatively good agreement between IOPst and IOPg,IOPst and IOPcc,IOPst and IOPnct,and CCTst and CCTp.The 95% limits of agreement were-2.62 to 4.91 mmHg,-3.04 mmHg to 6.22 mmHg,-2.29 to 4.44 mmHg and-15.57 to 13.59 μm.A positive correlation was found between IOPst and IOPg,IOPst and IOPnct,and CCTst and CCTp (r=0.774,0.791,0.981,P<0.05).Significant differences were found between IOPst and IOPg,IOPcc,IOPnct (t=-3.464,-3.922,-3.661,P<0.05),but there was no significant difference between CCTst and CCTp (t=0.778,P>0.05).Pearson analysis showed a high correlation between 1st A-time,Vout,HC-DA and CRF,CH (r=0.721,0.520,-0.545,-0.354,-0.596,-0.391,P<0.05).Conclusion In healthy eyes,the repeatability of some parameters using Scheimpflug Noncontact Tonometry (Corvis ST) is good.The agreement between Corvis ST and ORA and Pentacam was high.
What problem does this paper attempt to address?