Design of a portable SSVEP signal acquisition system

Yabin Dong,Lei Wang,Linlin Wang,Qian Li,Linyan Chai,Yan Wang
DOI: https://doi.org/10.3760/cma.j.issn.1673-4181.2019.03.007
2019-01-01
International Journal of Biomedical Engineering
Abstract:Objective To design a portable electroencephalography(EEG) acquisition system to acquire and analysis steady-state visual potentials (SSVEP). Methods The microprocessor MSP432P401 series MCU was used to control the high-performance integrated analog front end ADS1299 to realize the acquisition, amplification and analog-to-digital (AD) conversion of EEG signals. The digital EEG signal is sent to the host computer for processing by WIFI. Spontaneous EEG signals and steady-state visually evoked EEG signals from 3 healthy subjects were collected to verify system performance. Results The collected signal had a clear α-wave rhythm of closed-eye spontaneous EEG signals. The power spectrum density shows that the steady-state visually induced EEG signal frequency and harmonic frequency peak at the corresponding stimulation frequency, indicating that the system works normally and the performance is good. Conclusions The designed portable EEG acquisition system can accurately collect the spontaneous and induced EEG signals of the human body, which provides technical support for the clinical application of SSVEP technology.
What problem does this paper attempt to address?