Construction of bone tensile force monitoring system and its application in distraction osteogenesis

罗阳,王序全,张雪,王珏,侯天勇,张泽华,叶青,唐远林,陈卫军,府伟灵
DOI: https://doi.org/10.3760/cma.j.issn.1001-8050.2009.10.301
2009-01-01
Abstract:Objective To construct a kind of fixation and sensing system that can monitor the status of hone healing. Methods Solidworks2007 was adopted to design the novel full-ring fixator that was then compared with half-ring fixator in aspect of biedynamic stability. The whole system was assembled with minimized strain gauge drawing force and compression force transducer, data acquisition system and data analysis software. During distraction osteogenesis procedure, the change of tensile force of the tibia was monitored and X-ray photos taken to observe the bone callus state at different phases. Results The biodynamic test proved that the full-ring shackle external fixator could not only keep the advantage of easy fixation, but also enhance the ability against stress from lateral side and torque force compared with the half-ring fixator. Moreover, it guaranteed that three nods could endure the same stress during axial loading. During distraction osteogenesis, the full-ring fixation and sensing system could work real time assay and data analysis. X-ray and histomorphological results showed that the bone tensile force increased with distraction, with relatively poor bone healing but with short cost of time, which was in accordance with detection results of the bone stress. Conclusions We have successfully constructed a novel fullring fixator that can monitor the change of bone stress during bone healing, which provides a good basis for further study on mechanism of fracture healing.