Study of Influential Factors in Evaluating Tissue Stiffness by Virtual Touch Tissue Quantification

XUE Li-yun,WEN Jie-xian,LU Qing,YAN Cui-ju,HUANG Bei-jian,WANG Wen-ping
DOI: https://doi.org/10.3760/cma.j.issn.1004-4477.2012.06.022
2012-01-01
Abstract:Objective To investigate the influence of different depths and mediums on tissue stiffness measurement by virtual touch tissue quantification technology.Methods Four kinds of in vitro porcine tissue were recruited,three (fat,muscle and liver) were assigned as medium group and three (kidney,liver and muscle) as sample group.The shear wave velocity (Vs) of the samples was measured under different mediums and at different depths (2 cm,3 cm,4 cm and 5 cm),respectively.Results According to the depth of measurement,two groups of Vs were defined with one group at the depth of 2 cm and 3 cm while the other at 4 cm and 5 cm.The difference of Vs between the two groups was of statistical significance and with the increase of the depth Vs values decreased.Even at the same depth,the differences of Vs under different mediums were statistically significant.Conclusions Depths and mediums have great influence on evaluating elasticity measurements by virtual touch tissue quantification technology.
What problem does this paper attempt to address?