The Instant Biomechanical Stability of Anterior Lumbar Interbody Fusion

张继东,夏群,胡永成,苗军,种涛
DOI: https://doi.org/10.3760/cma.j.issn.0253-2352.2010.05.010
2010-01-01
Abstract:Objective To evaluate the instant biomechanical stability of anterior lumbar interbody fusion(ALIF)with a poly-ether-ether-ketone(PEEK)cage.Methods Six male human lumbar spines were collected; X-ray examination was performed to exclude pathology and obvious degenerative changes of the spines.Bone mineral densities of these spines were normal,identified by LUNAR DPX-L X-ray Bone Den-sitometer.L4-S2 segment were prepared and embedded to clamps.Preloads of 500 N was applied to speci-mens through longitudinal axis of the spine.Then the specimens were tested on the following 3 modes in flexion,extension and lateral bending: 1)intact lumbar spine; 2)stand alone ALIF on L5S1; 3)ALIF aug-mented with pedicle screw system.Results The range of motion(ROM)of stand alone ALIF decreased76.9%,66.6%,81.1% in flexion,extension and lateral bending respectively.The ROM of ALIF augmented with pedicle screw system further decreased 93.9%,90%,86.6% in flexion,extension and lateral bending respectively.The ROM of L4,5 increased significantly after stand alone ALIF alone and ALIF augmented with pedicle screw system in L5S1.Conclusion The instant stabilities of stand alone ALIF with PEEK cage are much better than intact lumbar spine.Pedicle screw system fixation can reinforce stability of ALIF signifi-cantly.