Optical Design and Simulation of an Integrated OCT and Video Rigid Laryngoscope

Jiajing Kang,Xinyu Li,Mingming Wan,Jiahui Wang,Shanshan Liang,Jun Zhang
DOI: https://doi.org/10.1142/s1793545820400027
IF: 2.396
2020-01-01
Journal of Innovative Optical Health Sciences
Abstract:An integrated optical coherence tomography (OCT) and video rigid laryngoscope have been designed to acquire surface and subsurface tissue images of larynx simultaneously. The dual-modality system that is based on a common-path design with components as few as possible effectively maintains the light transmittance without compromising the imaging quality. In this paper, the field of view (FOV) of the system can reach [Formula: see text] by use of a gradient index (GRIN) lens as the relay element and a four-lens group as the distal objective, respectively. The simulation showed that the modulation transfer function (MTF) value in each FOV of the rigid video endoscope at 160[Formula: see text]lp/mm is greater than 0.1 while the root mean square (RMS) radii of the OCT beam in the center and edge of the FOV are 14.948[Formula: see text][Formula: see text]m and 73.609[Formula: see text][Formula: see text]m, respectively. The resolutions of both OCT and video endoscope meet the requirement of clinical application. In addition, all the components of the system are spherical, therefore the system can be of low cost and easy to assemble.
What problem does this paper attempt to address?