Development of Microfluidic Platform to High-Throughput Quantify Single-Cell Intrinsic Bioelectrical Markers of Tumor Cell Lines, Subtypes and Patient Tumor Cells

Yi Zhang,Hongyan Liang,Huiwen Tan,Deyong Chen,Yixiang Wang,Ying Xu,Junbo Wang,Jian Chen
DOI: https://doi.org/10.1016/j.snb.2020.128231
2020-01-01
Abstract:Although intrinsic bioelectrical markers of single cells (e.g. specific membrane capacitance of C-sm, cytoplasm conductivity of sigma(cy) and cell diameter of D-c) play key roles in cell-type classification and cell-status evaluation, due to technical limitations, data from large populations of single cells were not available. This paper presents a microfluidic platform to high-throughput quantify single-cell intrinsic bioelectrical markers of tumor cell lines, subtypes and patient tumor cells, where an asymmetrical constriction channel defined by a major and two side constriction channels was used as the sensitive unit. The developed microfluidic platform was used to quantify similar to 100,000 single-cell C-sm, sigma(cy) and D-c from (I) eight tumor cell lines of A549, Hep G2, SW620, AGS, PANC-1, Hela, CAL 27 and HL-60 with cell-type classification rates of 87.6 % +/- 9.1 % obtained; (II) tumor cell subtypes of 293 T with and without the transfection of the CRISPR knockout library where different distributions of C-sm were located; (III) patient tumor cells with classification rates of 96.1 % +/- 3.8 % obtained compared with tumor cell lines. In conclusion, the microfluidic platform developed in this study can function as a high-throughput tool in the field of single-cell analysis.
What problem does this paper attempt to address?