Reversible Microfluidic Measurement with Zinc Tetraphenylporphyrin Aiming at Slipping NH3detection

Hao Zhou,Qi Yin,Zheng Yang,Yilin Song
DOI: https://doi.org/10.1049/mnl.2020.0135
2020-01-01
Micro & Nano Letters
Abstract:A microfluidic chip that can realise reversible and highly selective NH3 detection by separating NH3 from the complex environment is fabricated for slipping NH3 detection. The microfluidic chip consisting of the reaction unit, the gas-diffusion unit and detection unit applied the immobilised zinc tetraphenylporphyrin as the indicating material. The continuous spectrum change showed the light intensity decreased in the green range (500–550 nm). Based on the analysis of the spectrum change at 528 nm, a one-to-one relationship between the spectrum change and concentration was revealed. The reversibility of the detection process makes the detection advantageous. Also, the rapidity of the proposed method is proved by the time responses of the indicating and recovering processes. According to the 40 tests conducted for various concentrations, it was concluded that the results are stable and the entire system is reliable.