A Co-Axial Projection Surgical Navigation System for Breast Cancer Sentinel Lymph Node Mapping: System Design and Clinical Trial

Peng Liu,Pengfei Shao,Jianmin Ma,Min Xu,Chenmeng Li
DOI: https://doi.org/10.1117/12.2509852
2019-01-01
Abstract:Sentinel lymph node biopsy is important in the early stage breast tumor resection surgery. Its results will determine if the axillary lymph node dissection ( ALND) will be conducted afterwards. For locating sentinel lymph nodes, indocyanine green ( ICG) has been widely used with a near infrared ( NIR) camera to image its fluorescence. However, surgeons need to watch a screen beside the operating table to see the fluorescence, with their hands operating on the surgical site. We developed a navigation system that projects the invisible fluorescence back to the surgical site visibly in real-time. The system introduces a co-axial optics design to guarantee the projection accuracy. Phantom experiments are conducted to assess the projection resolution and accuracy of the system. Animal experiments with three mice show a good system performance and its preclinical feasibility. Furthermore, the system is tested in a clinical trial of ninety breast cancer patients in three hospitals in China. ICG and methylene blue ( MB) is subcutaneously injected separately into the areola at 3-4 points to get both fluorescent and visible contrast, for further comparison. The navigation process is compared with a commercialized NIR imaging system. The results show a 100% detection rate of sentinel lymph nodes and a good consistency with the methylene blue and the commercialized imaging device. The experiments demonstrate good clinical feasibility of the co-axial projection system.
What problem does this paper attempt to address?