A Reconstruction Method for Non-rigid Motion Compensation in Brain MRI

Feng Huang,Wei Lin,Chiel den Harder,Gabrielle Beck,Clemens Bos,George Randy Duensing,Arne Reykowski
2011-01-01
Abstract:Introduction Most existing motion compensation techniques for brain MRI [1-6] assume that head motion is rigid. In fact, many kinds of non-rigid motion such as eye movement (eye ball rolling), skin movement (frowning), and jaw movement (swallowing, yawning), can also cause serious spatially local artifacts in brain imaging. These inevitable problems have not been carefully addressed. To remove these artifacts, methods comprising data rejection and reconstruction with remaining unpolluted data can be used [4-6]. However, the reconstruction with partial data could result in potential artifacts, such as reduced signal to noise ratio (SNR) and loss of contrast. In this work, it is proposed to use the image reconstructed with all of the acquired kspace data, which is locally artifact corrupted but with high SNR, as a regularization term in reconstruction to achieve an image with low artifact level and high SNR.