Interlaboratory comparison of 25-hydroxyvitamin D assays: Vitamin D Standardization Program (VDSP) Intercomparison Study 2 — Part 1 liquid chromatography – tandem mass spectrometry (LC-MS/MS) assays — impact of 3-epi-25-hydroxyvitamin D3 on assay performance
Stephen A. Wise,Johanna E. Camara,Carolyn Q. Burdette,Grace Hahm,Federica Nalin,Adam J. Kuszak,Joyce Merkel,Ramón A. Durazo-Arvizu,Emma L. Williams,Andrew N. Hoofnagle,Fiona Ivison,Ralf Fischer,Jody M. W. van den Ouweland,Chung S. Ho,Emmett W. K. Law,Jean-Nicolas Simard,Renaud Gonthier,Brett Holmquist,Sarah Meadows,Lorna Cox,Kimberly Robyak,Michael H. Creer,Robert Fitzgerald,Michael W. Clarke,Norma Breen,Pierre Lukas,Étienne Cavalier,Christopher T. Sempos
DOI: https://doi.org/10.1007/s00216-021-03576-1
2021-08-25
Analytical and Bioanalytical Chemistry
Abstract:An interlaboratory comparison study was conducted by the Vitamin D Standardization Program (VDSP) to assess the performance of liquid chromatography – tandem mass spectrometry (LC-MS/MS) assays used for the determination of serum total 25-hydroxyvitamin D (25(OH)D), which is the sum of 25-hydroxyvitamin D2 (25(OH)D2) and 25-hydroxyvitamin D3 (25(OH)D3). A set of 50 single-donor samples was assigned target values for concentrations of 25(OH)D2, 25(OH)D3, 3-epi-25-hydroxyvitamin D3 (3-epi-25(OH)D3), and 24R,25-dihydroxyvitamin D3 (24R,25(OH)2D3) using isotope dilution liquid chromatography – tandem mass spectrometry (ID LC-MS/MS). VDSP Intercomparison Study 2 Part 1 includes results from 14 laboratories using 14 custom LC-MS/MS assays. Assay performance was evaluated using mean % bias compared to the assigned target values and using linear regression analysis of the test assay mean results and the target values. Only 53% of the LC-MS/MS assays met the VDSP criterion of mean % bias ≤ |±5%|. For the LC-MS/MS assays not meeting the ≤ |±5%| criterion, four assays had mean % bias of between 12 and 21%. Based on multivariable regression analysis using the concentrations of the four individual vitamin D metabolites in the 50 single-donor samples, the performance of several LC-MS/MS assays was found to be influenced by the presence of 3-epi-25(OH)D3. The results of this interlaboratory study represent the most comprehensive comparison of LC-MS/MS assay performance for serum total 25(OH)D and document the significant impact of the lack of separation of 3-epi-25(OH)D3 and 25(OH)D3 on assay performance, particularly with regard to mean % bias.
chemistry, analytical,biochemical research methods