Continuous measurement system of blood pressure based on A-type ultrasonic wave sensor

Hong-yang LI,Yi-bin LI,Shen-long LI,Ning DENG
DOI: https://doi.org/10.13873/J.1000-9787(2016)02-0087-04
2016-01-01
Abstract:A new continuous measurement system of blood pressure using A-type ultrasonic wave ranging method is designed. The system uses ultrasonic wave sensor to detect carotid artery,obtain carotid artery diameter waveform by recognizing the reflected ultrasonic wave from front and back walls of the carotid artery to measure distance. Then establish blood pressure calculation model based on relationship between artery diameter and blood pressure,so as to measure blood pressure. Experimental results show that compared with the standard blood pressure measurement method,the maximum standard deviation of systolic blood pressure is 4. 72 mmHg,the maximum standard deviation of diastolic blood pressure is 3. 17 mmHg,which coincides with the standard of not more than 8 mmHg made by AAMI.