Compare Wide-field Swept-Source Optical Coherence Tomography Angiography with Ultra-wide Field Color Fundus Photo and Fluorescein Angiography for Detecting Diabetic Retinopathy …

Ying Cui,Ying Zhu,Jay Wang,Yifan Lu,Rebecca Zeng,Raviv Katz,David M Wu,Demetrios G Vavvas,Deeba Husain,Joan W Miller,Leo A Kim,John B Miller
2020-01-01
Abstract:Purpose: To compare wide-field swept-source optical coherence tomography angiography (WF SS-OCTA) with ultra-wide field color fundus photo (UWF CFP) and fluorescein angiography (FA) for detecting diabetic retinopathy lesionsMethods: We conducted a prospective, observational study at Mass Eye and Ear from December 2018 to October 2019. Proliferative diabetic retinopathy (PDR), non-proliferative diabetic retinopathy (NPDR) and diabetic patients with no DR were included. All patients were imaged with a WF SS-OCTA with Montage 15mm× 9mm. UWF CFP and FA were taken by a 200 degree, single capture retinal imaging system. Images were independently evaluated by two graders for the presence or absence of DR lesions including microaneurysms, intraretinal microvascular abnormalities (IRMA), neovascularization elsewhere (NVE) and neovascularization of the optic disc (NVD), nonperfusion …