Image Reconstruction Using Multi-Energy System Matrices with a Scintillator-Based Gamma Camera for Nuclear Security Applications

Qi Liu,Yi Cheng,Yongliang Yang,Ying Peng,Hongyu Li,Yisheng Xiong,Tao Zhu
DOI: https://doi.org/10.1016/j.apradiso.2020.109217
IF: 1.787
2020-01-01
Applied Radiation and Isotopes
Abstract:The system response of a gamma camera is dependent on the photon energy, thus the energy-dependent response function needs to be considered to improve the quality and fidelity of reconstructed images for identifying radionuclides in security applications. In this study, two reconstruction strategies using the maximum-likelihood expectation maximization (MLEM) algorithm with the multi-energy system matrices calculated by Monte Carlo simulations are proposed. The difference between the two is in data acquisition; one uses the sum of all events into a single projection image while the other sorts them into separate energy windows. Various radiation images of gamma-ray sources were simulated with a Monte Carlo code, and an actual image was acquired with a gamma camera. Both simulation and experiment results demonstrated the feasibility of the presented multi energy reconstruction strategies in the detection of orphan sources.
What problem does this paper attempt to address?