Phrenic Nerve Injury During Right Inferior Pulmonary Vein Ablation with the Second-Generation Cryoballoon:A Report of 2 Cases and Literature Review

Xianxian Zhou,Xuping Li,Hui Tian,Zhihong Wu,Zhenjiang Liu,Hui Yang,Qiming Liu,Shenghua Zhou
DOI: https://doi.org/10.11817/j.issn.1672-7347.2020.190002
2020-01-01
Abstract:The performance of the second-generation cryoballoon has been greatly improved compared to the first-generation.But it has not reduced the incidence of phrenic nerve injury, and even increased the incidence of this complication during ablation of the right inferior pulmonary vein. Two cases of phrenic nerve injury occurred when right inferior pulmonary vein was ablated with the second-generation cryoballoon in Department of Cardiovascular Medicine, Second Xiangya Hospital, Central South University. Based on literature review, the possible causes for phrenic nerve injury in our report were summarized to provide helpful information for operators aiming to reduce the occurrence of this complication related to the second-generation cryoballoon.
What problem does this paper attempt to address?