A Strip Test for the Optical Determination of Influenza Virus H3 Subtype Using Gold Nanoparticle Coated Polystyrene Latex Microspheres.

Xiao Liu,Jifei Yang,Qingmei Li,Yinbiao Wang,Yanhong Wang,Ge Li,Jianzhou Shi,Peiyang Ding,Junqing Guo,Ruiguang Deng,Gaiping Zhang
DOI: https://doi.org/10.1007/s00604-020-04255-1
2020-01-01
Mikrochimica Acta
Abstract:A strip test is described for the optical determination of influenza virus H3 subtype. It utilizes gold nanoparticle (AuNP) coated polystyrene latex microspheres (PS) as the label and a sandwich format. The AuNP and PS particles were linked using monoclonal antibodies against influenza virus as the bridge. Under the optimal conditions, the visual detection limit of the AuNP-PS-based strip test was as low as 1/16 hemagglutination unit (HAU). It was 64 times higher than that of 10 nm (4 HAU) AuNP-based strip tests. Quantitative analysis showed that the detection limit of the AuNP-PS-based strip is 0.016 HAU. The AuNP-PS-based strip test showed no cross-reactivity to the other subtypes (H1, H5, H7, or H9) of influenza viruses.
What problem does this paper attempt to address?