Cavity ring-down spectroscopy measurements of ambient NO<sub>3</sub> and N<sub>2</sub>O<sub>5</sub><SUP>?</SUP>

Hao Wu,Jian Chen,An-wen Liu,Shui-ming Hu,Jing-song Zhang
DOI: https://doi.org/10.1063/1674-0068/cjcp1910173
IF: 1.09
2020-01-01
Chinese Journal of Chemical Physics
Abstract:NO<INF>3</INF> and N<INF>2</INF>O<INF>5</INF> are important participants in nocturnal atmospheric chemical processes, and their concentrations are of great significance in the study of the mechanism of nocturnal atmospheric chemical reactions. A two-channel diode laser based cavity ring-down spectroscopy (CRDS) instrument was developed to monitor the concentrations of NO<INF><INF>3</INF></INF> and N<INF>2</INF>O<INF>5</INF> in the atmosphere. The effective absorption length ratio and the total loss coefficient of the instrument were calibrated using laboratory standard samples. The effective absorption cross section of NO<INF>3</INF> at 662 nm was derived. A detection sensitivity of 1.1 pptv NO<INF>3</INF> in air was obtained at a time resolution of 1 s. N<INF>2</INF>O<INF>5</INF>was converted to NO<INF>3</INF> and detected online in the second CRDS channel. The instrument was used to monitor the concentrations of NO<INF>3</INF> and N<INF>2</INF>O<INF>5</INF> in the atmosphere of winter in Hefei in real time. By comparing the concentration changes of pollutants such as nitrogen oxides, ozone, PM<INF>2.5</INF> in a rapid air cleaning process, the factors affecting the concentrations of NO<INF>3</INF> and N<INF>2</INF>O<INF>5</INF> in the atmosphere were discussed.
What problem does this paper attempt to address?