Improving axial resolution of Bessel beam light-sheet fluorescence microscopy by photobleaching imprinting

Bo Xiong,Xiaofei Han,Jiamin Wu,Hao Xie,Qionghai Dai
DOI: https://doi.org/10.1364/OE.388808
IF: 3.8
2020-01-01
Optics Express
Abstract:Light-sheet microscopy has been widely used in high-speed fluorescence imaging with low phototoxicity, while the trade-off between the field-of-view and optical sectioning capability limits its application in large-scale imaging. Although Bessel beam light-sheet microscopy greatly enhances the light-sheet length with the self-healing ability, it suffers from the strong side-lobe effect. To solve these problems, we introduce the photobleaching imprinting technique in Bessel beam light-sheet microscopy. By extracting the non-linear photobleaching-induced fluorescence decay, we get rid of the large concentric side lobe structures of the Bessel beam to achieve uniform isotropic resolution across a large field-of-view for large-scale fluorescence imaging. Both numerical simulations and experimental results on various samples are demonstrated to show our enhanced resolution and contrast over traditional Bessel-beam light-sheet microscopy. (C) 2020 Optical Society of America under the terms of the OSA Open Access Publishing Agreement
What problem does this paper attempt to address?