Wearable Airflow Sensor For Nasal Symmetric Evaluation And Respiration Monitoring

Tao Jiang,Junwen Zhong,Jiaming Liang,Yichuan Wu,Zehui Li,Xiaohao Wang,Liwei Lin,Xiang Qian
DOI: https://doi.org/10.1109/SENSORS43011.2019.8956504
2019-01-01
Abstract:Respiration is an important physiological function in which air is moved into and out of the lungs. This paper designed a flexible patch for continuous detections of respiration, especially for evaluating nasal symmetry. The monitoring system comprises an integrated hot-film and two time-of-flight flow sensors on a flexible polyimide (PI) substrate. The 5 mu m-thick polyimide substrate is flexible to construct wearable devices while the hot-wire flow sensor can monitor the respiration conditions and states, such as different breathing patterns of mouth or nose modes, normal or deep forms, and happy or nervous signs. Moreover, the power consumption of the system is low for possible long-term monitoring applications of sleep and medical diagnosis. The two time-of-flight sensors are aligned symmetrically on different sides of the hot-film heating element for differential signal readings, which used asymmetrical thermal distribution to reveal breathing state differences of two nostrils and they don't require additional energy since they are located near the heating element. The system serves as a noninvasive wearable device to monitor respiratory airflow while avoiding uncomfortable nasal cannula. Various testing results show the feasibility and effectiveness of the proposed system.
What problem does this paper attempt to address?