An Immunoassay Based on Lab-on-a-chip for Simultaneous and Sensitive Detection of Clenbuterol and Ractopamine

Qi Wang,Jinqi Deng,Yiping Chen,Yunjing Luo,Xingyu Jiang
DOI: https://doi.org/10.1016/j.cclet.2020.01.028
IF: 9.1
2020-01-01
Chinese Chemical Letters
Abstract:Both clenbuterol (CLB) and ractopamine (RAC) are beta-adrenergic agonists. After long-term excessive intake, there will be adverse reactions such as headache, chest tightness, limb numbness, and serious lifethreatening. Simultaneous detection of CLB and RAC in related samples is of great importance for human health. In this work, we outline a microfluidics-based indirect competitive immunoassay (MICI) system that can sensitively detect residual CLB and RAC in pork, swine blood and swine urine. The rapid detection of multiple samples can be achieved in one chip, which greatly improves the detection efficiency. This method has good stability and reproducibility and the microfluidic chips are easy to manufacture. The linear ranges for CLB and RAC detection by MICI are 0.1-2.5 ng/mL and 0.1-5 ng/mL, and the limits of detection (LODs) are 0.094 ng/mL and 0.091 ng/mL, respectively. This straightforward and portable immunoassay system provides a good platform for rapid detection of harmful substances in food samples. (C) 2020 Chinese Chemical Society and Institute of Materia Medica, Chinese Academy of Medical Sciences. Published by Elsevier B.V. All rights reserved.
What problem does this paper attempt to address?