Droplet Digital PCR Enabled by Microfluidic Impact Printing for Absolute Gene Quantification

Yang Pan,Tuo Ma,Qi Meng,Yuxin Mao,Kaiqin Chu,Yongfan Men,Tingrui Pan,Baoqing Li,Jiaru Chu
DOI: https://doi.org/10.1016/j.talanta.2019.120680
IF: 6.1
2020-01-01
Talanta
Abstract:Digital PCR enabled high-sensitivity and quantitative measurements of rare biological variants. A new digital droplet-enabled PCR technology was introduced in this paper, which partitioned genetic targets into a planar nanoliter droplet array by using a microfluidic impact printer (MIP) with a disposable microfluidic chip. The accuracy of this MIP-enabled PCR technology was verified by detecting a series of concentration gradients of GAPDH gene across spanning four orders of magnitude (from 0.464 copies/μL to 464 copies/μL). Furthermore, this technology was applied to detect the expressions of p53 gene in colon cancer tissues and adjacent nontumorous tissues, from which the copies of the nucleic acids could be absolute-quantitatively determined. The outcomes were consistent with the results of using the conventional real-time PCR, demonstrating a great potential of the MIP-enabled digital PCR in detecting gene expression in clinical samples.
What problem does this paper attempt to address?