On-site Testing of Multiple Drugs of Abuse in Urine by a Miniature Dual-Lit Mass Spectrometer.

Manqing Kang,Wanru Zhang,Linpei Dong,Xinxin Ren,Yin Zhu,Zhenhua Wang,Lijun Liang,Jinfeng Xue,Yunfeng Zhang,Wenpeng Zhang,Zheng Ouyang
DOI: https://doi.org/10.1016/j.aca.2019.12.028
IF: 6.911
2020-01-01
Analytica Chimica Acta
Abstract:There is an increasing need for rapid and on-site detection of emerging drugs of abuse. In this work, we developed a method using a miniature dual-LIT (linear ion trap) mass spectrometer recently developed with comprehensive tandem mass spectrometry analysis capability, for qualitative and quantitative analysis of multiple drugs of abuse. Paper-capillary spray cartridges were used with related workflow established to simplify overall analysis procedure. Quantitation of ketamine and methamphetamine was achieved by beam-type collision-induced dissociation on the miniature dual-LIT mass spectrometer and a linear concentration range of 100-5000 ng/mL was obtained. The system has been applied in analysis of real urine samples from individuals addicted to morphine and methamphetamine use. The changes of the ratio of cocaine to its metabolite benzoylecgonine were also explored to estimate the time of cocaine intaking. (C) 2019 Elsevier B.V. All rights reserved.
What problem does this paper attempt to address?