Ultrasound and Contrast-enhanced Ultrasound Findings after Percutaneous Irreversible Electroporation of Hepatic Malignant Tumors.

Weilu Chai,Liting Xie,Qiyu Zhao,Chao Cheng,Guo Tian,Tian'an Jiang,Pingping Wu
DOI: https://doi.org/10.1016/j.ultrasmedbio.2019.12.012
IF: 3.694
2020-01-01
Ultrasound in Medicine & Biology
Abstract:The aim of this study was to describe ultrasound (US) and contrast-enhanced ultrasound (CEUS) findings immediately and 1 d after percutaneous irreversible electroporation (IRE) of hepatic malignant tumors. Immediately after IRE, the ablation zone was shown to be a gradually expanding hypo-echoic area around the electrodes. The microcirculation of the ablation zone was markedly reduced on CEUS (before vs. immediately after, p < 0.001), and the macrocirculation within the ablation zone was preserved. At 1 d after IRE, the ablation zones lost their hypo-echogenicity to become iso-echoic or hyper-echoic (before vs. 1 d after, p = 0.004; immediately after vs. 1 d after, p = 0.002). At this time, further elimination of microcirculation was confirmed on CEUS (before vs. 1 d after, p < 0.001; immediately after vs. 1 d after, p = 0.003). The size of the ablation zone, which measured by US, was strongly correlated with that measured by CEUS (length: r: = 0.929, width: r = 0.940, p < 0.001), was significantly enlarged immediately after IRE and shrunk 1 d after IRE.
What problem does this paper attempt to address?