An Ultrasensitive Electrochemiluminescence Resonance Energy Transfer Biosensor for Divalent Mercury Monitoring

Shu-Ping Cao,Hui-Min Hu,Ru-Ping Liang,Jian-Ding Qiu
DOI: https://doi.org/10.1016/j.jelechem.2019.113494
IF: 4.598
2020-01-01
Journal of Electroanalytical Chemistry
Abstract:An electrochemiluminescence (ECL) sensing approach was designed for sensitive Hg2+ detection based on oligonucleotide hosted silver cluster (DNA-Ag NCs) as effective unit for switching the ECL signal of graphite-like carbon nitride nanosheets (g-C3N4 NSs). In this system, g-C3N4 NSs were first immobilized on the glassy carbon electrode surface followed by the chitosan coating and probe DNA assembly. The g-C3N4 NSs exhibited stable ECL intensity which was regarded as first signal switch "on" state. DNA-Ag NCs can be captured onto the g-C3N4 NSs coated electrode surface via DNA hybridization, leading to the ECL signal quenching of g-C3N4 NSs attributed to the ECL resonance energy transfer (ECL-RET) between the luminophore donor g-C3N4 NSs and the acceptor DNAAg NCs, turning to signal "off" state. While in the presence of Hg2+, the formation of T-Hg2+-T coordination between DNA-Ag NCs or DNA-Ag NCs themselves would hinder DNA-Ag NCs from embellishing on the electrode surface, resulting in the recovery of the g-C3N4 NSs ECL intensity corresponding to the last signal switch "on" state. Based on this, the biosensor can be demonstrated to quantitatively analyze Hg2+ with a detection of limit of 5 pM. The proposed ECL platform has been utilized to sensitive and selective monitoring Hg2+ in environmental water samples.
What problem does this paper attempt to address?