Ultrasensitive Quantitation of Imidacloprid in Vegetables by Colloidal Gold and Time-Resolved Fluorescent Nanobead Traced Lateral Flow Immunoassays

Guiyu Tan,Yajie Zhao,Mian Wang,Xiaojiao Chen,Baomin Wang,Qing X. Li
DOI: https://doi.org/10.1016/j.foodchem.2019.126055
IF: 8.8
2019-01-01
Food Chemistry
Abstract:Imidacloprid is a highly effective insecticide, but its potential hazards to the environment and ecosystems have limited its use in many regions. A fast and sensitive analytical method would aid monitoring imidacloprid residues. The monoclonal antibody 4D9, colloidal gold and time-resolved fluorescent nanobeads (CGN and TRFN, respectively) were applied to develop a lateral flow immunoassay (LFIA) for imidacloprid detection in the present work. Both the optimized TRFN-LFIA and CGN-LFIA had a similar limit of detection of approximately 0.02 ng/mL. Use of the portable optical scanner with the TRFN-LFIA or CGN-LFIA could quantitate concentrations of imidacloprid in vegetable as sensitively as the enzyme-linked immunosorbent assay (ELISA). The LFIAs quantitative results of imidacloprid in commercial Chinese leeks were verified by the liquid chromatography-mass spectrometry with a R-2 of 0.91. The LFIAs are portable and simple and thus could fully replace the ELISA for onsite quantitation of imidacloprid residues in vegetables.
What problem does this paper attempt to address?