Thiol-functionalized Reduced Graphene Oxide As Self-Assembled Ion-to-electron Transducer for Durable Solid-Contact Ion-Selective Electrodes.

Yueling Liu,Yunzhong Liu,Zhen Meng,Yu Qin,Dechen Jiang,Kai Xi,Ping Wang
DOI: https://doi.org/10.1016/j.talanta.2019.120374
IF: 6.1
2019-01-01
Talanta
Abstract:Thiol-functionalized reduced graphene oxide (TRGO) as a novel ion-to-electron transducing layer is firstly employed to develop durable solid-contact ion-selective electrodes (SC-ISEs) in this work. The performance of the sensors is evaluated by determining K+ and NO3- as an example of cation and anion, The covalent linkage of TRGO at golden electrode surface generates a stable transducing layer. No water films are observed in the proposed TRGO-based potassium (K+-TRGO-ISEs) and nitrate (NO3--TRGO-ISEs) selective SC-ISEs. The resultant electrodes exhibit Nemstian responses (60.0 +/- 0.4 mV/decade for K+-TRGO-ISEs and -60.0 +/- 0.5 mV/decade for NO3--TRGO-ISEs), low detection limits (2.5 x 10(-6)M for K+-TRGO-ISEs and 4.0 x 10(-6)M for NO3--TRGO-ISEs) and good selectivity behavior. More importantly, the TRGO-based SC-ISEs display a much longer lifetime of 2 weeks than that of reduced graphene oxide-based SC-ISEs in continuous flowing solutions using a longer peristaltic pump. These improvements push TRGO a general and reliable transducer for the development of durable SC-ISEs.
What problem does this paper attempt to address?