Electron Modulated Arc Therapy (EMAT) Using Photon MLC for Postmastectomy Chest Wall Treatment I: Monte Carlo-based Dosimetric Characterizations.

Chaoqiong Ma,David Parsons,Mingli Chen,Steve Jiang,Qing Hou,Xuejun Gu,Weiguo Lu
DOI: https://doi.org/10.1016/j.ejmp.2019.10.018
IF: 3.119
2019-01-01
Physica Medica
Abstract:PURPOSE:To study the dosimetric properties of electron arc beams delivered by photon-beam multi-leaf collimators (pMLC) in electron modulated arc therapy (EMAT) for postmastectomy chest wall treatments.METHODS:Using the Monte Carlo method, we simulated a 2100EX Varian linear accelerator and verified the beam models in a water tank. Dosimetric characterizations were performed on cylindrical water phantoms of elliptical bases with various field sizes, arc ranges and source-to-surface distances (SSDs) for 6, 9 and 12 MeV beam energy.RESULTS:The arc beam has a higher bremsstrahlung dose than the static beam at the isocenter due to crossfire, but choosing a field size greater than 5 cm effectively reduces the bremsstrahlung dose. The depths of the 90% maximum dose located at 1.7, 2.8 and 4.1 cm for 6, 9 and 12 MeV, respectively, are similar to those of the static beams and independent of the field size and arc range.CONCLUSION:Based on the study, we recommend using the 5 cm field width for electron arc beams considering both bremsstrahlung dose at the isocenter and the arc profile penumbra. To ensure sufficient PTV edge coverage, we recommend a field length extension of at least 4 cm from PTV's edge for all beam energies and an arc extension of around 7°, 5°, and 5° for beam energies 6, 9, and 12 MeV, respectively. These dosimetric characterizations are the basis of pMLC-delivered EMAT treatment planning for postmastectomy chest wall patients.
What problem does this paper attempt to address?