Single-level Cervical Arthroplasty with ProDisc-C Artificial Disc: 10-Year Follow-Up Results in One Centre.

Zhao Yanbin,Zhou Feifei,Sun Yu,Pan Shengfa
DOI: https://doi.org/10.1007/s00586-019-06110-1
IF: 2.721
2019-01-01
European Spine Journal
Abstract:The aim of this study was to evaluate the long-term clinical and radiographic outcomes of cervical arthroplasty using the ProDisc-C prosthesis. Clinical and radiographic evaluations, including dynamic flexion–extension lateral images, were performed at baseline and at 10-year follow-up. Twenty-seven patients who had single-level ProDisc-C arthroplasty were followed up for a mean period of 123 months. The range of motion at the operated level was 8.9° ± 3.9° at baseline and 6.6° ± 3.5° at final follow-up. Twenty of 27 levels (74%) developed heterotopic ossification. According to McAfee’s classification, one level was classified as grade I, four levels were classified as grade II, 12 levels were classified as grade III and three levels were classified as grade IV. Three patients developed recurrent cervical radiculopathy or myelopathy due to adjacent segment disease and received the reoperations. The reoperations included two cases of cervical arthroplasty at adjacent segments and one case of cervical laminoplasty. ProDisc-C arthroplasty had acceptable clinical and radiographic results at 10-year follow-up. Heterotopic ossification was common after ProDisc-C arthroplasty, which decreased the range of motion. These slides can be retrieved under Electronic Supplementary Material.
What problem does this paper attempt to address?