A Sensitive Determination of Albumin in Urine by Molecularly Imprinted Electrochemical Biosensor Based on Dual-Signal Strategy

Guihua Zhang,Ying Yu,Manli Guo,Bixia Lin,Li Zhang
DOI: https://doi.org/10.1016/j.snb.2019.03.042
2019-01-01
Abstract:The detection of albumin in urine is of great significance for the early diagnosis and treatment of diseases. Herein, a molecularly imprinted electrochemical biosensor was designed for the detection of human serum albumin (HSA) in urine based on dual-signal strategy. The gold electrode (GE) was orderly modified by Au nanoparticles (AuNPs), polythionine-methylene blue (PTH-MB) and molecularly imprinted polymers (MIPs). The MIPs were synthesized through electro-polymerization, which utilized HSA as the template, and employed o-phenylenediamine (o-PD) and hydroquinone (HQ) as the functional monomers. The biosensor showed dual-signals at 0.20 V and -0.22 V corresponding to Fe(CN)(6)(3-/4-) and PTH-MB, respectively. When HSA was captured on the MIPs, the dual-signal simultaneously decreased. Both of the current signals of the substrate probe PTH-MB (Delta I-substrate) and the solution probe [Fe(CN)(6)](3-/4-)(Delta I-probe) were converted before and after the incubation of target HSA. The sum of the signal conversions Delta I-D (Delta I-substrate + Delta I-probe) was utilized to quantify the target amount. Under the optimal conditions, the proposed biosensor exhibited a linear regression equation in the concentration range from 1.0x10(-10) to 1.0x10(-4) g L-1, with a low detection limit of 3.0x10(-11) g L-1 (S/N = 3). Furthermore, the prepared biosensor was employed for the determination of albumin in urine with satisfied selectivity and reproducibility.
What problem does this paper attempt to address?