Optimization Of Parabolic Cell For Gas Raman Analysis

Anlan Yu,Duluo Zuo,Xinbing Wang
DOI: https://doi.org/10.1002/jrs.5564
IF: 2.727
2019-01-01
Journal of Raman Spectroscopy
Abstract:Raman spectroscopy is a widely practicable technique in gas analysis, but the application in simultaneous detection of multiple-trace gas is still a challenge for its weak signal level. To extend the application of Raman gas analysis to detection of trace gas constituents, a closed parabolic sample cell proposed in our previous work is optimized by optical analysis and experimental verification. The optical analysis shows that for a parabolic reflector with rim diameter 6p (p, semi-latus rectum), the collecting solid angle can be as large as 90% of full space (4 pi) and the power-collecting efficiency can be as high as 80%. The variation of the Raman signal level with the position of collecting aperture and the parameter p of parabolic reflector was verified by the experimental results. At the optimized structure parameters, we get a signal-to-background ratio of 122 and a signal-to-noise ratio of 305 in the Raman spectra of ambient air with exposure time of 1 s. From the results of standard analytic gaseous samples, the limits of detection of 33 ppm-bar for H-2, 38 ppm-bar for CO, 19 ppm-bar for CH4, 27 ppm-bar for C2H6, 23 ppm-bar for C2H4, and 20 ppm-bar for C2H2 were obtained with exposure time of 200 s. These results indicate that the parabolic sample cell is a competitive Raman collector, which showing good balance in signal enhancement and background level.