S-Sequence Encoded Multiplane Wave Imaging: Phantom and In-Vivo Validation

Yinran Chen,Xianquan Shi,Linxue Qian,Jianwen Luo
DOI: https://doi.org/10.1109/ultsym.2018.8579979
2018-01-01
Abstract:Plane wave imaging with spatial coherent compounding (PWC)shows potential in providing ultrafast imaging, at the expense of low signal-to-noise ratio (SNR)due to lack of focusing in transmit. In order to increase the SNR without sacrificing the frame rate, Hadamard-encoded (HD-encoded)multiplane wave imaging (MWI)is proposed by transmitting several steered plane waves simultaneously, with the pulses encoded by the row entries of a Hadamard matrix. Then, the backscattered echoes are decoded by using the column entries of the same Hadamard matrix. However, zero-degree plane wave is always missing due to the constraint of Hadamard matrix. Moreover, pulse inversion operation in HD-encoded MWI leads to the risk of introducing extra harmonic artifacts in conventional B-mode imaging. In order to mitigate these problems, S-sequence (SS)is used in combination with MWI in this study, as an alternative encoding-decoding method. Phantom and preliminary in-vivo experiments show that the SS-encoded MWI achieves higher frame rate and moderate imaging quality when compared with related HD-encoded MWI and PWC.
What problem does this paper attempt to address?