Measurement of Proton Beam Generated Β+ Radioactivity by Use of All-digital PET Detectors

Min Gao,Nicola D'Ascenzo,Chien-Min Kao,Yuexuan Hua,Xudong Lyu,Ing-Tsung Hsiao,Chung-Yi Li,Hsien-Hsin Chen,Hong,Tzu-Chen Yen,Qingguo Xie
DOI: https://doi.org/10.1109/nssmic.2017.8532734
2017-01-01
Abstract:One key issue in proton beam therapy is to ensure precise delivery of the concentrated radiation dose to the target. A promising method that is under active study and development is the use of position emission tomography (PET) for range verification or monitoring. In this work, we test the use of a digital PET system developed on the basis of a Plug-and-Imaging architecture for a proton therapy system. We analyze technical features of the proposed PET technology and experimentally demonstrate the possibility to imaging positron decays in a phantom immediately after proton irradiation. The experiment results agree with the expectation according to the physical properties of proton beam interaction with hydrocarbons.
What problem does this paper attempt to address?