Study and Design on Anti-Coincidence High-Energy Γ Photon Spectrometer

孙长春,李家才,姚淑德,张亚伟,张少平
DOI: https://doi.org/10.3969/j.issn.0258-0934.2004.04.014
2004-01-01
Abstract:In this paper, using EGS4 program code we have studied and simulated the transport of high energy γ with 6.6 MeV in the CsI crystal, decided the size of CsI crystal to detect such energyγphoton. According to the need of experiment and result of the simulation, finally we designed the anti-coincidence high-energyγphoton spectrometer. The measurements to this spectrometer by ~(137)Cs show that the efficiency of photoelectric peak at 0.662 MeV is about 74%. And the calculation shows that it is about 37.12% and 23.5% separately at 3.3 MeV, 6.6 MeV.