Experimental Study of Intravascular Ultrasonic Ablation Angioplasty for Acute Canine Arterial Thrombi

郭大乔,符伟国,蒋俊豪,杨珏,史振宇,王玉琦
DOI: https://doi.org/10.3969/j.issn.0253-9934.2003.08.008
2003-01-01
Abstract:Objective To evaluate the efficacy and safety of intravascular ultrasound angioplasty in the ablation of acute canine arterial thrombi. Methods Bilateral femoral acute thrombotic occlusions were created in 12 mongrel dogs with a thrombus more than 5 cm long. In all the dogs, the right femoral artery served as the test artery and the ultrasound wire was introduced to the site of thrombus with application of ultrasound ablation for 2 minutes, while the left side served as control which was mechanically penetrated with the ultrasound device without ablation. The change of blood flow and the degree of arterial wall injury at the site of thrombi as well as the distal embolization in each dog were measured by flowmeter, angiography, and pathologic examination. Results After sonication, the blood flow distal to the thrombotic canine femoral artery was (28.9±4.2)ml/min, significantly higher than (9.6±5.0)ml/min in the control ( P 0.01). The TIMI grade index was 2.67±0.49 and 1.08±0.90 in the ultrasound group and the control group respectively( P 0.01). However, the arterial wall injury index was noted to be 1.83±0.39 in the tested arteries compared with 1.67±0.49 in the control arteries ( P 0.05).No distal embolization was found in the ultrasound ablation group. Conclusion Intravascular ultrasonic angioplasty has a potential to ablate acute arterial thrombi effectively and safely.
What problem does this paper attempt to address?