Long-term Safety and Efficacy of Combined Percutaneous LAA and PFO/ASD Closure: a Single-Center Experience (LAAC Combined PFO/ASD Closure)

Jiangtao Yu,Xiaoxia Liu,Junling Zhou,Xin Xue,Manuela Muenzel,P. Christian Schulze,Sven Moebius-Winkler,Thorsten Keil,Zhaohui Meng,Shaoyong Tang
DOI: https://doi.org/10.1080/17434440.2019.1604216
IF: 3.439
2019-01-01
Expert Review of Medical Devices
Abstract:OBJECTIVES:To report long-term safety and efficacy of combined percutaneous LAA and PFO/ASD closure.METHODS:A retrospective study of 370 consecutive patients undergoing LAAC procedures using the Watchman (WM) device. Data were compared between 330 cases only with LAAC procedure (Group I) and 25/5 (PFO/ASD) cases with sequential procedures of LAAC and PFO/ASD closure (Group II).RESULTS:Compared to Group I, Group II had more males (86.7% vs. 65.8%, p < 0.05) and a higher rate of stroke (33.3% vs. 10.6%, p < 0.01), but there were no statistical differences in the remaining patient characteristics. During the follow-up period, there were no significant differences between the two groups in embolism events (6.1% vs. 0%, p = 0.39), device related thrombus (5.8% vs 3.3%, p = 1.0), major bleeding (9.4% vs. 6.7%, p = 1.0) and cardiac death (3.6% vs. 0%, p = 0.61). The observed rate of all thromboembolic events by Kaplan-Meier analysis was decreased by 39.9% and 100% and the observed annual rate of bleeding was reduced by 32.9% and 57.6% in Group I and Group II, respectively.CONCLUSIONS:LAAC combined with PFO/ASD closure might be an ideal choice to prevent stroke and other thrombotic complications in patients with both NVAF and PFO/ASD.
What problem does this paper attempt to address?