Application of Different Near-infrared Spectrometers in Tea Quality Inspection

NIU Zhi-You,LIU Mei-ying,LIN Xin
DOI: https://doi.org/10.3969/j.issn.0439-8114.2009.10.074
2009-01-01
Abstract:Two different types of NIRSystems 6500 and InfraXact Lab near-infrared spectrometer were used to establish calibration models for the tea polyphenol and caffeine of tea.The optimal spectral data pretreatments were confirmed with the modified partial least squares calibration technology for each spectrometer.The optimal calibration models of tea polyphenol and caffeine in tea were established.The tea polyphenol target value f were 96.07% and 95.96% of two different types nearinfrared spectrometer,respectively.The caffeine target value f were 95.32% and 91.47%,respectively.The optimal calibra-tion models were tested by validation sets,the result of RPD was greater than 3.The results showed that the calibration effects of 2 models were good,and there were high prediction accuracy.Both of two types of pretreatments can be used in actual testing,but the calibration effect and stability of the NIRSystems 6500 near-infrared spectrometer was better than the type of InfraXact Lab.