Quantitative Analysis of Microcirculation in Renal Solid Lesions with Gamma Curve Fitting Technique of Contrast-Enhanced Ultrasound

李萍,李凤华,方华,顾莉红,杜晶,黄翼然
2009-01-01
Abstract:Objective To evaluate the role of gamma curve fitting technique of contrast-enhanced ultrasound in quantitative analysis of microcirculation in renal solid lesions.Methods A total of fifty patients with renal parenchyma solid lesions were performed contrast-enhanced ultrasound.The images were analysed by computer with gamma fitting analysis of contrastenhanced ultrasonic system.The quantitative parameters were obtained by the time-intensity curves,such as ascending slope (a_3),descending slope(a_2),arrival time(AT),time to peak intensity(TTP),basic intensity(BI),peak intensity, amplification(AMP),area under the curve(AUC),mean transit time(MTT) and perfusion index(PI).The parameters were compared between renal malignant and benign solid lesions.Results Fast-in and fast-out was the main perfusion mode in renal malignant tumors while slow-in and slow-out was found in renal angiomyolipoma(AML).The perfusion modes in renal malignant tumors and renal AML were fast-in and fast-out in 28 cases and O case,fast-in and slow-out in 4 cases and 1 case,slow-in and fast-out in 5 cases and 1 case,and slow-in and slow-out in 1 case and 10 cases,respectively.There were significant differences in the quantitative parameters such as a_3,AUC and PI between renal malignant tumors and renal AML obtained by the time-intensity curves(P0.05).Conclusion Gamma fitting analysis of contrast-enhanced ultrasound system can provide quantitative information of microcirculation of renal tumors,which helps to differentiate benign renal tumors from malignant ones.