Determination of Tyrosine and Tyrosine metabolites in Biological Sample by Liquid Chromatography-Tandem Mass Spectrometry

Lei FENG,Yaqian MA,Ting TAN,Aiping YAN,Lan GUO,Yi-qun WAN
DOI: https://doi.org/10.3969/j.issn.1006-0464.2014.04.012
2014-01-01
Abstract:A simple,selective,rapid and reliable method for the simultaneous determination of 3,4-di-hydroxyphenylalanine (DOPA),tyrosine (Tyr),4-hydroxyphenyllactic acid (PHPLA)and 4-hydroxyphen-ethylamine (TA)in human serum by high-performance liquid chromatography (HPLC)coupled with tan-dem mass spectrometry (MS/MS)was proposed for the first time.Under optimized conditions,all com-pounds were resolved within 6 min.The calibration curves were indicative of good linearity (correlation co-efficients >0.998 9),and the limits of detections and quantifications for Tyr and its metabolites ranged from 0.000 5 to 0.20 mg·L-1 and 0.001 to 0.40 mg·L-1 ,respectively.Recoveries of Tyr and its metabo-lites in breast cancer patient were in the range of 85.2%~114.0%,with RSDs of 0.7%~8.5%.The pres-ent method was successfully applied to the determination of DOPA,Tyr,PHPLA and TA in sera from breast cancer patients.
What problem does this paper attempt to address?