Study on Opening Electrical Impedance Tomography Principles and Imaging Experiment

刘斌,何为,张晓菊,何传红
DOI: https://doi.org/10.3969/j.issn.1003-8868.2009.04.001
2009-01-01
Abstract:Objective To develop a high-precision measuring system based on opening electrical impedance tomography(OEIT) for the survey of physical model and clinical application.Methods The fixed electrode array that can conquer the problem of unequal distance and inconvenient arrangement of the electrode in CEIT was used in the OEIT.The direct digit synthesis(DDS) was employed by measuring system to produce the exciting signal,and FPGA and ADC were developed to realize digital phase-sensitive wave detecting and obtain the voltage amplitude and phase of the electrodes.Finally,the conductivity image distribution was given by computation according to numerical algorithm.Results The signal to noise ratio(SNR) of the measurement system was up to 104 dB,and can distinguish two objects with conductivity of 400 mS/m and 200 mS/m respectively under 300 mS/m electric conductivity.Conclusion The measurement system can satisfy with the requirement of SNR to OEIT.The two objects experiment in water-tank and agar model shows that OEIT has good resolution and positioning accuracy,can be applied to distinguish the small difference in hypo-conductivity structure.