OPTIMIZING INSTRUMENTAL CONDITIONS FOR METHYLMERCURY ANALYSE BY USING AUTOMATED METHYLMERCURY ANALYTICAL SYSTEM

陈来国,田文娟,莫测辉,许振成,彭晓春,方建德
IF: 4.236
2011-01-01
Environmental Chemistry
Abstract:It is possible to analyze MeHg rapidly and precisely with the development of automated methylmercury(MeHg) analytical system,in which the optimization of instrumental parameters and sample pretreatment procedure are the keys.In this study,we optimized instrumental analysis conditions to obtain the best results for the separation and run time of ethylated Hg0,MeHg,and Hg2+.The optimum condition is the following: the flow rates of purging gas nitrogen and carrier gas argon were 344 mL·min-1 and 34 mL·min-1,respectively.And the GC column temperature was about 40℃.For ethylation reagent(1% NaBEt4 solution),an optimum volume of 40 μL was selected when the content of MeHg ranged from 0 to 100 pg.Finally,this optimized method was used to analyze target chemicals in fish samples digested separately by KOH/H2O and KOH/CH3OH solutions.The results showed good accuracy and precision that meet the requirement for analysis of MeHg in fish.