Rapid Detection of HPV Using Leaky Surface Acoustic Wave Peptide Nucleic Acid Biosensor and Regeneration of Ability

WANG Yun-xia,ZHANG Li-qun,LUO Yang,DING Yi,SHI Jian-feng,CAO Liang,CHEN Ming,FU Wei-ling
DOI: https://doi.org/10.3321/j.issn:1005-4529.2009.14.010
2009-01-01
Abstract:OBJECTIVE To detect HPV using leaky surface acoustic wave(LSAW) peptide nucleic acid(PNA) biosensor,and study the regeneration of the sensor.METHODS The same concentration probes were immobilized and hybridized with complementary targets,then repeated the experiments for 20 times.Before each reaction,Piranha solution,1 mol/L HCl and 1 mol/L NaOH were used respectively to wash the reaction area and to explore the regeneration of the sensor.RESULTS After five regenerating cycles,the biosensor could retain 89.60% of the original response;the coefficient of variance(CV) was 4.24%.After ten regenerating cycles the biosensor retained 71.56% of the original response and the CV was 10.14%.After 10 times the detection ability of the biosensor decreased quickly.CONCLUSIONS LSAW-bis-PNA biosensor has good reproducibility.Piranha solution,1 mol/L HCl and 1 mol/L NaOH can be used to wash everything covered on the reaction area,including probe,target and hybridized nucleic acid molecules.
What problem does this paper attempt to address?