Respiratory Amplitude Measurement in Target Localization for Focused Ultrasound Surgery

高爽,陈亚珠,高巍
DOI: https://doi.org/10.3969/j.issn.1002-3208.2002.02.009
2002-01-01
Abstract:Focused Ultrasound Surgery (FUS) is a promising minimally invasive surgery technique. Target localization is one of the most important aspects that determine if FUS will be successful or not. Respiration movement is the primary factor affecting target localization while the patient is in a quiet state. In order to analyze the relationship between respiratory amplitude and target motion, a respiratory amplitude measurement system was developed. This system acquires respiratory amplitude data by measuring the variation of circumference of thorax during breathing. This system mainly consists of respiratory amplitude transducer, signal conditioning and amplifying circuits, data acquisition board and computer. For the sake of verifying the feasibility of the developed system, another system, which acquires respiratory amplitude data by calculating the variation of position of diaphragm in medical ultrasonic images during breathing, was also designed. These two systems were used to acquire synchronously respiratory amplitude data while the subjects were in quiet breathing state. After the two sets of data were preprocessed, the cross\|correlation processing was done and the cross\|correlation coefficient was 0\^9765. Meanwhile, the calculated maximum movement of the diaphragm was 13\^88mm, which was in the normal range. The above mentioned results indicate that the developed respiratory measurement system is feasible.