Rapid Determination of Trace Iodide in Solutions by UV Spectrophotometry

ZHANG Hui-fang,LIU Hai-ning,GUO Min,YE Xiu-shen,LI Quan,WU Zhi-jian
2011-01-01
Abstract:A rapid UV spectrophotometric method was established to determine trace iodide content in solutions.The linear range of iodide solution at 226 nm is 0.005~0.200 mmol/L.The linear fitting equation is A=12.413c+0.039 with a correlation coefficient of 0.9991.The detection limit is 9.909×10-5 mmol/L.The RSD(n=6) of the analysis method is 0~0.789%.The recovery for oilfield water sample is 98.65~100.00%.Temperature,solution pH,and the coexisting K+,Na+,Br-,Cl-and SO2-4 ions do not affect the detection in general.The coexisting CO2-3 and HCO-3 ions have much interference in the determination,which can be eliminated by acidification.The coexisting Mg2+ and Ca2+ ions have an obvious effect on the detection,which can be calibrated through the linear relationships between the absorbance and Mg2+ or Ca2+concentrations in solutions with a fixed I-content.The method could be used to determine trace iodide content in oilfield water.
What problem does this paper attempt to address?