Experimental Research on Zn Ion Selective Electrode and Measurement of Minim Element

胡卫军,王平
DOI: https://doi.org/10.3969/j.issn.1000-9787.2008.05.018
2008-01-01
Abstract:The free Zn2+ is difficult to preserve in low concentration.The existing form of Zn in the solution system is discussed and the free Zn2+ with stable and low concentration are achieved by means of calculation of Visual MINTEQ software based on the complex equilibrium of Zn.After the calibration to the buffer solution with the Zn ion selective electrode(Zn-ISE),it seems that Zn-ISE can measure normally when the concentration of the free Zn2+ is higher than 10-8mol/L.The measurement limit of Zn-ISE can be extended from 10-5mol/L to 10-8mol/L.
What problem does this paper attempt to address?