Abstract:Objective An animal experiment was carried out to complete proximal anastomosis of vein graft and ascending a orta in CABG with Symmetry Aortic Proximal Connector in dog model. Methods The proximal anastomosis of the OPCABG completed with the Symmetry Aortic Proximal Connector were assigned as the experiment group(N=10),while the proximal anastomosis manually sutured as the control group(N=10).The time of completing the anastomosis and the flow through the vein grafts were recorded,The leakage of the anastomosis was identified at different blood pressure levels,which were raised by venous bolus of neosynephrine. Results The anastomotic time consuming of th e experiment group was 98.30 ± 9.63 seconds,while the control group was 200.50±7.54 seconds(P0.01). The blood flow through vein graft in experiment group was 12.07± 0.60ml/min,and 11.96±0.80 ml/min in control group (P0.05).The blood pressure threshold of anastomosis bleeding were180±7.0mmHg and 160 ± 8.0mmHg in experiment group and control group(P0.01). respectively. Conclusion Symmetry Aortic Proximal Connector is a quick,simple,effective and reliable device for proximal anastomosis of coronary artery bypass surgery.