Integration of Chip-Scale SERF Atomic Magnetometers for Magnetoencephalography Measurement

Guoliang Li,Jintang Shang,Yu Ji,Lin Lu,Zhihua Pan
DOI: https://doi.org/10.1109/ectc.2018.00124
2018-01-01
Abstract:This paper presents a novel scheme on the integration of chip-scale SERF magnetometer using a vertical-cavity surface-emitting laser(VCSEL) as pumping and detecting laser. The fabrication and integration process of the whole system based on stacking PCB boards consist of VCSEL, optical components, rubidium vapor cell, photo-detector, three axis Helmholtz coils. By calculation and measurement, it was found that the module is designed with a minimum output power of 0.15mW and a competitive volume of 1.4×1.4×2.5cm 3 . Experiments show that the sensitivity of chip-scale SERF magnetometer based on VCSEL is around 1.87pT/Hz 1/2 in magnetic shield environment, the integration and miniaturization design process of the magnetometer can be empolyed for magnetoencephalography(MEG) measurement.
What problem does this paper attempt to address?