Abstract:The reliability and validity of 3D virtual Stent numerical simulation was investigated by using particle image velocimetry (PIV) experiments. A real 3D stent was designed by Computer Aided Design (CAD), and then fabricated by a 3D rapid prototyping print machine (EDEN250). In order to keep the same flow fields in the amplified model and numerical simulation, the flow velocity in the model was determined by the Reynolds similarity criterion. For easy measurements and after many experiments, the dimensions of the stent were magnified 10 times. The printed stent was put in a transparent tube of diameter of 30 mm for decreasing the effects of laser's secondary refringence near the wall of the tube. The intravascular flow field was measured by particle image velocimetry (PIV). The experiment results verified the effectiveness of the numerical simulations of the stent.