An Intracranial Implantable Ultrasound Device for Seizure Mapping

Weiwei Shao,Yuanmin Miao,Zhangjian Li,Zhile Han,Jibing Wu,Dan Li,Hongtao Ma,Yaoyao Cui
DOI: https://doi.org/10.1109/ultsym.2017.8092594
2017-01-01
Abstract:For drug-resistant epilepsy patients, epilepsy surgery is the only option. In clinic, this is done with subdural electrode array with a disappointing low spatial resolution of ~ 1cm. In order to improve clinical cure rate, ultrasound imaging technique can be cooperated with Electrocorticography to monitor the maps of seizure. We propose to develop the implanted single ultrasound transducers 3mm×1.5mm×0.8mm to satisfy clinical requirement. For the single transducer recording, the laws of epilepsy corresponding with ultrasonic signals need to be rediscovered. In order to verify our proposals, the phantom experiment is designed to simulate the epileptogenic zone. The experiment result shows the power spectral density is in direct proportion to simulation Cerebral Blood Volume increased. This device makes high resolution 3D seizure maps possibly to guide the epilepsy surgery.