Particle identification technique using grid ionization chamber at China Spallation Neutron Source

Yingtan Zhao,Ruirui Fan,Qingmin Zhang,Yuefeng He,Zhijia Sun,Liang Zhou,Jingyu Tang,Huaiyong Bai,Haoyu Jiang,Guohui Zhang
DOI: https://doi.org/10.1088/1748-0221/12/11/P11001
2017-01-01
Journal of Instrumentation
Abstract:Back-n white neutron source at China SpallationNeutron Source facility requires a detector with charged-particle identification ability for measuring nuclear cross sections. Therefore, a grid ionization chamber (GIC) with an adjustable distance between the cathode and grid was designed and built. Moreover, a corresponding particle identification algorithm is proposed in order to use the signals of the anode (Q(A)) and cathode (Q(C)), or the signal (Q(A)) of the anode and the time difference (Delta t) between the cathode and anode signals, based on the interactive behaviors of charged particles and the GIC's working principle. First, the GIC performance was measured using a Am-241 alpha source for preliminary performance testing, which provided an energy resolution of 5.25%. Its good particle identification performance was demonstrated by identifying H-3 and He-4 from neutron reactions with Li-6 using a 2D Q(C)-Q(A) plot and t-Q(A) plot. Furthermore, H-3 and He-4 can be completely distinguished from the updated 2D Q(C)-Q(A) plot and t-Q(A) plot, in which H-3 with a small incident angle is selected. Finally, it was concluded that our GIC with the proposed particle identification algorithm could identify charged particles effectively for the future measurement of nuclear cross sections.
What problem does this paper attempt to address?